ISO 80601-2-84:2020
Medical electrical equipment - Part 2-84: Particular requirements for the basic safety and essential performance of ventilators for the emergency medical services environmentstandard by International Organization for Standardization, 07/01/2020
In stock
ISO 80601-2-84:2020 applies to the basic safety and essential performance of an EMS ventilator in combination with its accessories, hereafter also referred to as ME equipment:
intended for patients who need differing levels of support from artificial ventilation including ventilator-dependent patients; intended to be operated by a healthcare professional operator; intended for use in the EMS environment; and intended for invasive or non-invasive ventilation.NOTE 1 An EMS ventilator can also be used for transport within a professional healthcare facility.
* An EMS ventilator is not considered to utilize a physiologic closed loop-control system unless it uses a physiological patient variable to adjust the ventilation therapy settings.
This document is also applicable to those accessories intended by their manufacturer to be connected to the ventilator breathing system, or to an EMS ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the EMS ventilator.
NOTE 2 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005, 7.2.13 and 8.4.1.
NOTE 3 Additional information can be found in IEC 60601-1:2005+AMD1:2012, 4.2.
This document does not specify the requirements for the following:
ventilators or accessories intended for ventilator-dependent patients in critical care applications, which are given in ISO 80601-2-12. ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601-2-72[3]. ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601213[4]. ventilators or accessories intended for ventilatory support equipment (intended only to augment the ventilation of spontaneously breathing patients), which are given in ISO 80601-2-79[5] and ISO 80601-2-80[6] [1]. obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601270[7]. operator-powered resuscitators, which are given in ISO 106514[8]. gas-powered emergency resuscitators, which are given in ISO 106515[9]. continuous positive airway pressure (CPAP) ME equipment . highfrequency jet ventilators (HFJVs), which are given in ISO 80601-2-87[11]. highfrequency oscillatory ventilators (HFOVs)[10], which are given in ISO 80601-2-87[11].NOTE 4 An EMS ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilation-modes.
cuirass or ironlung ventilators.[1] ISO 80601-2-79 and ISO 80601-2-80 replace ISO 10651-6, which has been withdrawn.